Filtered by vendor Ishekar
Subscribe
Total
7 CVE
| CVE | Vendors | Products | Updated | CVSS v2 | CVSS v3 |
|---|---|---|---|---|---|
| CVE-2017-10724 | 1 Ishekar | 2 Endoscope Camera, Endoscope Camera Firmware | 2024-11-21 | 6.5 MEDIUM | 8.8 HIGH |
|
Recently it was discovered as a part of the research on IoT devices in the most recent firmware for Shekar Endoscope that an attacker connected to the device Wi-Fi SSID can exploit a memory corruption issue and execute remote code on the device. This device acts as an Endoscope camera that allows its users to use it in various industrial systems and settings, car garages, and also in some cases in the medical clinics to get access to areas that are difficult for a human being to reach. Any breac ...
Show More |
|||||
| CVE-2017-10723 | 1 Ishekar | 2 Endoscope Camera, Endoscope Camera Firmware | 2024-11-21 | 6.5 MEDIUM | 8.8 HIGH |
|
Recently it was discovered as a part of the research on IoT devices in the most recent firmware for Shekar Endoscope that an attacker connected to the device Wi-Fi SSID can exploit a memory corruption issue and execute remote code on the device. This device acts as an Endoscope camera that allows its users to use it in various industrial systems and settings, car garages, and also in some cases in the medical clinics to get access to areas that are difficult for a human being to reach. Any breac ...
Show More |
|||||
| CVE-2017-10722 | 1 Ishekar | 2 Endoscope Camera, Endoscope Camera Firmware | 2024-11-21 | 4.6 MEDIUM | 7.8 HIGH |
|
Recently it was discovered as a part of the research on IoT devices in the most recent firmware for Shekar Endoscope that the desktop application used to connect to the device suffers from a stack overflow if more than 26 characters are passed to it as the Wi-Fi password. This application is installed on the device and an attacker who can provide the right payload can execute code on the user's system directly. Any breach of this system can allow an attacker to get access to all the data that th ...
Show More |
|||||
| CVE-2017-10721 | 1 Ishekar | 2 Endoscope Camera, Endoscope Camera Firmware | 2024-11-21 | 4.0 MEDIUM | 6.5 MEDIUM |
|
Recently it was discovered as a part of the research on IoT devices in the most recent firmware for Shekar Endoscope that the device has Telnet functionality enabled by default. This device acts as an Endoscope camera that allows its users to use it in various industrial systems and settings, car garages, and also in some cases in the medical clinics to get access to areas that are difficult for a human being to reach. Any breach of this system can allow an attacker to get access to video feed a ...
Show More |
|||||
| CVE-2017-10720 | 1 Ishekar | 2 Endoscope Camera, Endoscope Camera Firmware | 2024-11-21 | 4.6 MEDIUM | 7.8 HIGH |
|
Recently it was discovered as a part of the research on IoT devices in the most recent firmware for Shekar Endoscope that the desktop application used to connect to the device suffers from a stack overflow if more than 26 characters are passed to it as the Wi-Fi name. This application is installed on the device and an attacker who can provide the right payload can execute code on the user's system directly. Any breach of this system can allow an attacker to get access to all the data that the us ...
Show More |
|||||
| CVE-2017-10719 | 1 Ishekar | 2 Endoscope Camera, Endoscope Camera Firmware | 2024-11-21 | 4.0 MEDIUM | 6.5 MEDIUM |
|
Recently it was discovered as a part of the research on IoT devices in the most recent firmware for Shekar Endoscope that the device has default Wi-Fi credentials that are exactly the same for every device. This device acts as an Endoscope camera that allows its users to use it in various industrial systems and settings, car garages, and also in some cases in the medical clinics to get access to areas that are difficult for a human being to reach. Any breach of this system can allow an attacker ...
Show More |
|||||
| CVE-2017-10718 | 1 Ishekar | 2 Endoscope Camera, Endoscope Camera Firmware | 2024-11-21 | 4.0 MEDIUM | 6.5 MEDIUM |
|
Recently it was discovered as a part of the research on IoT devices in the most recent firmware for Shekar Endoscope that any malicious user connecting to the device can change the default SSID and password thereby denying the owner an access to his/her own device. This device acts as an Endoscope camera that allows its users to use it in various industrial systems and settings, car garages, and also in some cases in the medical clinics to get access to areas that are difficult for a human being ...
Show More |
|||||