Vulnerabilities (CVE)

Filtered by vendor Philips
Filtered by product Intellivue Mx40 Firmware
Angry Yack Logo
Total 2 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2017-9658 1 Philips 2 Intellivue Mx40, Intellivue Mx40 Firmware 2024-11-21 6.1 MEDIUM 6.5 MEDIUM
Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitorin ...

Show More

CVE-2017-9657 1 Philips 2 Intellivue Mx40, Intellivue Mx40 Firmware 2024-11-21 3.3 LOW 6.5 MEDIUM
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient ex ...

Show More