Vulnerabilities (CVE)

Filtered by vendor Medtronic
Filtered by product 2090 Carelink Programmer Firmware
Angry Yack Logo
Total 3 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2018-5448 1 Medtronic 2 2090 Carelink Programmer, 2090 Carelink Programmer Firmware 2025-05-22 2.7 LOW 4.8 MEDIUM
Medtronic 2090 CareLink Programmer’s software deployment network contains a directory traversal vulnerability that could allow an attacker to read files on the system.
CVE-2018-5446 1 Medtronic 2 2090 Carelink Programmer, 2090 Carelink Programmer Firmware 2025-05-22 2.1 LOW 4.9 MEDIUM
Medtronic 2090 CareLink Programmer uses a per-product username and password that is stored in a recoverable format.
CVE-2018-10596 1 Medtronic 2 2090 Carelink Programmer, 2090 Carelink Programmer Firmware 2025-05-22 5.2 MEDIUM 7.1 HIGH
Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-s ...

Show More