vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
| Link | Resource |
|---|---|
| https://www.radiometer.com/myradiometer |
No configuration.
18 Dec 2025, 15:08
| Type | Values Removed | Values Added |
|---|---|---|
| New CVE |
Published : 2025-12-17 13:15
Updated : 2025-12-18 15:08
NVD link : CVE-2025-14097
Mitre link : CVE-2025-14097
CVE.ORG link : CVE-2025-14097
JSON object : View
No product.
Improper Authentication