vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
| Link | Resource |
|---|---|
| https://www.radiometer.com/myradiometer |
No configuration.
18 Dec 2025, 15:08
| Type | Values Removed | Values Added |
|---|---|---|
| New CVE |
Published : 2025-12-17 13:15
Updated : 2025-12-18 15:08
NVD link : CVE-2025-14096
Mitre link : CVE-2025-14096
CVE.ORG link : CVE-2025-14096
JSON object : View
No product.